Ultrasonic transducer array: PHC51 and PHC103V and PHL125, etc
Ultrasonic transducer accessories PHC51 array


Product name | convex array |
Product model | PHC51 |
Applicable OEM model | C5-1 |
Frequency | 1-5MHz |
Service category | Ultrasonic transducer accessories customization |
Warranty period | 1 year |
Delivery time: Arrange delivery as soon as the order is placed. If there is a large demand or special requirements, it will be determined according to the actual situation
Ultrasonic transducer accessories PHC103V array


Product name | Intracavity array |
Product model | PHC103V |
Applicable OEM model | C10-3V |
Frequency | 3-10MHz |
Service category | Ultrasonic transducer accessories customization |
Warranty period | 1 year |
Delivery time: Arrange delivery as soon as the order is placed. If there is a large demand or special requirements, it will be determined according to the actual situation
Ultrasonic transducer accessories PHL125 array


Product name | Linear array |
Product model | PHL125 |
Applicable OEM model | L12-5 |
Frequency | 5-12MHz |
Service category | Ultrasonic transducer accessories customization |
Warranty period | 1 year |
Delivery time: Arrange delivery as soon as the order is placed. If there is a large demand or special requirements, it will be determined according to the actual situation
How to identify ultrasonic transducer fault in preliminary?
Malfunctioning sound lens: Bubbles in the sound lens can cause partial black shadows on ultrasonic images; however, pressing down hard on the shadowed area may make it disappear. The damaged of acoustic lens will cause the coupling agent to penetrate into the crystal layer.
Sound head fault: The sound head fault is when the array element (crystal) has some sort of damage, and will appear as a dark channel, blood flow flower, or if it's concentrated in the middle then it will affect normal use.
Shell malfunction: The breaking of the shell will allow the coupling agent to penetrate into the probe, causing oxidation and corrosion of the sound head crystal.
Sheath fault: sheath is the protective layer of the cable, if it break there is a danger of damage to the cables.
Cable fault: The cable is the carrier connecting the sound head and the host system. The fault of the cable will cause the probe to appear dark channel, interference and ghosting.
Circuit fault: will lead to probe error, flaring, no recognition, double image, etc.
Oil sac fault: The oil sac be damaged can result in the leakage of oil, which will cause a black image to form locally.
Three-dimensional/Four-dimensional malfunction: Shows as three-dimensional/four dimensional not working (no image), motor does not work.